AdvantEDGE LD500 Local Display and PA500 Profibus Gateway0 pages
AdvantEDGE LD500 Local Display
Functions and Features
•
•
•
•
•
•
•
•
•
•
Connects to any AdvantEDGE system.
Convenient 1/4 DIN enclosure
Display sensor and temperature data
Menu Functions: Calibration, Configuration and Diagnostics
Digital Protocol Options for remote measurement, calibrations,
configurations, and diagnostics.
One 4 to 20 mA loop output (Standard)
Two 4 to 20 mA loop outputs (Optional)
Two form C relays for alarm, control, and wash (Optional)
PID for pH
7 Conduit knockouts for easy cabling
AdvantEDGE PA500 Profibus Gateway
PA500 Profibus Gateway Adapter connects
up to 30 Modbus Communications Modules
to a Profibus DP (V0 or V1) network with
M12 connector. Full-featured measurement,
calibration, configuration and diagnostics.
24 VDC power.
Mounting Options
• Conduit Box, NEMA 4X (Standard)
• 1/2 DIN NEMA 4X Panel Mount (Optional)
Interface
• Display: 2 line by 16 character LCD
• Backlight: high contrast green
• Keypad: 4 directional arrows, MENU, ESCAPE, and ENTER
Enviromental Conditions
•
•
•
•
Ambient Operating Temperature Range: -20 to 65 C
Maximum Relative Humidity: 95% non-condensing
Meets CE requirements for heavy industrial use.
Nema 4X housing.
AdvantEDGE Sensor Connections
Supports 1 to 255 Modbus RTU AdvantEDGE sensors of any type.
t
• Supported Sensors: pH, ORP, Two Electrode
Conductivity, Two electrode resistivity,
Toroidal Conductivity, Dissolved Oxygen,
Ozone, Turbidity, and Free Chlorine.
4 to 20 mA Current Loop Options
•
•
•
•
•
One Isolated Current Loop (Standard)
Two Isolated Current Loops (Optional)
Assignable Parameters
PID function when pH parameter selected
Loop power provided by LD500
Relays/Optional
• Two 1 Amp form C relays with NO and NC contacts.
• Assignable Parameters
• Configured for Alarm, Control, & Timer Functions
Digital Communication Options to Host Computer
Network
•
•
•
•
•
•
•
•
MODBUS RTU
Device Net
Profibus DP/V1
RS-232 ASCII
RS-232 DF1
RS485 ACSII
CANOpen
Others: Consult Factory
Power Supply
Input:t
10 to 30 VDC; 100 to 240 VAC optional
Power: t 2W at 25 C
03-17-2011 Rev 0.7