Heated Viscometer0 pages
Tel: +44(0) 1903 731 470 Kittiwake Developments Ltd.,3-6 Thorgate Road, Littlehampton,
West Sussex, BN17 7LU, United Kingdom. Fax: +44(0) 1903 731 480 E-mail: sales@kittiwake.com
1. Unheated with automatic viscosity correction to 40Cfor lube oils. 2.
nnH
nnHeated to 40аC for distillate fuels and lubricating oils.
nnEATED V ISCOMETER (
nnFG-K1-200-KW)
nnKnowledge of fuel oil viscosity is important for verifying thecorrect grade of fuel is delivered, for calculating combustion
performance, plus adjusting of fuel handling and injection
systems. Viscosity of lube oil is the most important property
of oil. Lube oil of the correct viscosity will provide optimum
film strength in bearing clearances with minimum friction
losses and leakage. The Kittiwake Fuel and Lube Oil Viscometer is extremelyrugged and ideal for harsh industrial applications. The
Viscometer is suitable for use on oils from a wide variety of
applications including diesel engines, gas and aviation
turbines, gear boxes, hydraulics and fuel oils.The
Kittiwake Viscometer can be used in 3 ways: 3. Heated to 50C for residual fuel oils.Readings are uniquely provided directly in centistokes. FEATURES
Monitors changes in lubricating oil viscosity,preventing costly engine and machinery failures.Verifies correct fuel grade or blend has beendelivered.Checks that fuel viscosity is acceptable forstorage, pumping and purifiers.Estimates the combustion performance (CCAI) offuel oil.Correct density from 50аC. to kg/m3@ 15C invacuo. >
nnSupplied with an extensive instrucution manual andTraining CDfor the use of the equipment together withinterpretation of results. nn
SPECIFICATIONS
WEIGHTS, DIMENSIONS, POWER
nnRange: 20-810cSt at 50аC (ISO Fuel Grades RMA10 to RML55) 20-810cSt at 40C (lubricating oils SAE5 through SAE50) >
nnTest Time:Heating from 25аC 10 mins. Viscosity at 40C unheated 3
mins. Repeat test 30 secs maximum >
Viscosity at 50аC or (heated) Viscosity at 40C (unheated, corrected to 40а heated)
Viscosity at 100C (calculated)
Calculated Aromaticity index (CCAI)
Density correlation from 50аC to 15C in vacuo
Variable viscosity Index (for unheated mode) >
SPARES
nnitem part no.Spare Seals, Fuses and Balls FG-K10503-KW Viscometer Power supply AS-K3-013-KW Viscometer End Cap AS-K11097 >
nnAccuracy:Typically within +/-3% (15-450cSt) Viscometer End Plug AS-K11098 Viscometer 24v Power Lead EL-K10450 >
nnCorrelation:ASTM D445/IP71 >
nnApplication:Fuel Oils and Lube Oils Power: 110/240 VAC user selected. Dimensions: 25 x 13 x 9 cm Weight: 10 kg >
CK50086 Issue 5 size="-2">