Model HG 800 pages
Information to
Model HG 80
Autoclave,
Chamber volume:
80 Litre
Chamber dimensions (mm):
364 Ǿ x 730 H
External dimensions (mm): 455B x 691T x 1030H
Working temperature:
105° C – 135° C
Max. Chamber pressure:
2,5 bar
Power:
3,7 KW
Power supply:
230 Volt/50Hz
Net weight:
73kg
Capacity baskets
3
HG 80
Description and technical information:
Contained in the base price:
Fully automatic microprocessor controlled
12 installed programs
Timer-controlled night program - overnight sterilization does not heat up the laboratory
AGAR program with thermostatic heating for AGAR Production
Liquid program with removal temperature of 95°
C
Small feed water consumption
Electronic pressure switch, pressure and temperature control
Exhaust bottle for condensation ‚ this prevents fogging and unpleasant odours
Drain valve for easy cleaning of sterilization chamber
Bottom plate for baskets and cans
Air-cooling by pressing a bottom usable
Lid-opening and lid-closing by pressing a bottom
On wheels - cleaning is possible even under the autoclave
Safety:
Safety valve
Pressure control of max. pressure
Limiter for max. chamber temperature
Safety sensor for heating elements
Approval by TÜV Bayern e.V.
CE-mark
Chamber and basics:
Lid with fast locking device and pressure are made from stainless steel (SUS 304). For easy cleaning
the chamber is polished. The housing is powder coated and also easy to clean. additionally heat
protection under the lid handle. The main switch is built-in and is protected by unintentional using.
Two recessed handles facilitate easy manipulation.
The feed-water-level is controlled and low water level is indicated. For protection of heating elements
the autoclave cannot be started without enough water. An optional feed water connection
automatically controls the water level. The exhaust steam condenser cools steam and condensate.
This prevents fogging of instruments and unpleasant odours.
Loading high is only 750mm without edges and eyelet at which baskets can hang remain.
The central quick lock can be actuated with one finger touch. An optional thermo lock with sterilization
time control prevent hazards to the laboratory staff by delayed boiling.
V:\HMC-Europe\Beschreibungstexte\HG\Datenblätter\Datenblätter englisch\Datasheet HG 80.doc2
"