Xenotest Alpha+ Brochure0 pages
Utility Requirements
Electrical 230 V ±10 %, 50/60 Hz
(1P,N,PE) AC or (2P,PE) AC | CEE (32 A, 3-pin. 6h)
Amperage 16 A
Maximum power consumption approx. 5 kVA
Cooling air requirement for xenon lamp 200 m3/h
Cooling air requirement for test chamber 100 m3/h
Water consumption for spray system 0.7 l/min
Water consumption for humidity max. 0.033 l/min
Sample Capacity
Sample holders 11*
Sample dimensions L x W (max.) 135 x 46 mm
Exposure Area 1320 cm2
*without additional sensor (22 samples during turning mode)
Physical Specifications
Width x Depth x Height 900 x 780 x 1800 mm
Weight approx. 280 kg
Atlas Material Testing Technology GmbH
(p) +49.60 51.707.140
(f) +49.60 51.707.149
Atlas Material Testing Technology LLC
(p) +1.773.327.4520
(f) +1.773.327.5787
www.atlas-mts.com
We reserve the right to make technical changes
to the instruments and systems.
©2006 Atlas Material Testing Technology GmbH
All rights reserved. Printed in Germany.
BV Pub. No. 56352317
US Pub. No. 2025
Xenotest® Alpha+ Features
Air-cooled xenon lamp providing a maximum of 2.2 KW 1
XENOSENSIV sensor to measure and control irradiance
from 300 to 400 nm and the Black Standard Temperature
at sample level 1
Temperature control selectable either by test chamber
temperature (up to 70° C) or both test chamber temperature
and Black Standard Temperature simultaneously (up to 130° C) 1
Air volume control to influence the temperature difference
between test chamber and Black Standard Temperature 1
Ultrasonic humidification system 1
Specimen spray system 1
Microprocessor control 1
Parameter check 1
User guided operation by color graphic display 1
Touch screen and I/O board using
optical fiber waveguide technology 1
Data output via memory card or RS232 / USB interface 1
Instrument-internal memory chip to store instrument data 1
Thermoprinter 1
XenoCal BST Black Standard Thermometer 1
XenoCal WST White Standard Thermometer 1
XenoCal BB 300–400 Irradiance Sensor 1
XenoCal WB 300–800 Irradiance Sensor 1
XenoCal NB 340 Irradiance Sensor 1
1Standard 1Optional
Xenotest® Alpha+ Specifications
Irradiance range: 300–400 nm at sample level (in W/m2)
Filter system Turning Mode Non-turning Mode
XENOCHROME 300 21–114 W/m2 40–220 W/m2
XENOCHROME 320 21–93 W/m2 40–180 W/m2
10 window glass 21–57 W/m2 40–110 W/m2
Absorption filter lantern with:
6 IR +1 UV 21–72 W/m2 40–140 W/m2
7 IR segments 21–72 W/m2 40–140 W/m2
4 IR +3 window segments 21–72 W/m2 40–140 W/m2
Daylight extended IR 21–65 W/m2 40–125 W/m2
IR absorption filter system 16H 21–72 W/m2 40–140 W/m2
Temperature and Humidity Range
Test chamber temperature: 30° C to 70° C*
Black Standard Temperature: 40° C to 130° C*
Relative humidity: 10 to 95%*
*Depending on the selected filter combination and irradiance
as well as the ambient laboratory conditions
Experience. The Atlas Difference.
Light Exposure
and Weathering
Testing Instrument
XENOTEST®
ALPHA+