Tabletop Microscope, TM30300 pages
TM3030 specification
I Specifications
Items | Description |
Magnification | 15 to SO.OOOx (digital zoom: 2x. 4x) |
Observation condition | 5kV/15kV/EDX |
Observation modo | Standard mode Charge-up reduction mode |
Image mode | COMPO/Shadow 1 /Shadow 2/TOPO |
Sample stage traverse | X: ±17.5mm, Y: ±17.5mm |
Maximum sample size | 70mm in diameter |
Maximum sample height | 50mm |
Electron gun | Pre-cerrtered cartridge filament |
Signal detection system | High-Sensitivity semiconductor 4-segment BSE detector |
Auto image adjustment function | Auto start, Auto focus, Auto brightness/contrasl |
Frame memory | 640 x 480 pixels, 1,280 x 960 pixels |
Image data memory | HDD of PC and other removal media |
Image format | BMP, TIFF, JPEG |
Data display | Micron marker, micron value, date and time, Image number and comments, Image mode, Observation condition, D* (Distance), Observation mode |
Evacuation system (vacuum pump) | Turbomolecular pump: 301/s x 1 unit, Diaphragm pump: 1m3/h x 1 unit |
Operation help functions | Raster rotation, Magnification preset (two steps) Image shift (±50um@D*=4.5mm) |
Safety device | Over-current protection function, built-in ELCB |
*D (Distance) Is defined as the *
I Required PC specifications
Items | Description |
OS | Windows7 Professional (32/64bit version) |
CPU | lntel®Core™ i5-2520M (Equivalent or higher) |
Memory | 2GB minimum |
Display resolution | 1,280x800 pixels or 1,366x768 pixels |
Display size | 15-Type display |
Interface connector | Installing USB2.0 and PC-card slot (IEEE1394 (6pin) for Oxford EDX is indispensable.) |
Memory device | With HDD DVD-ROM Drive |
Other | More than 100MB of free space in HDD is required |
*An associated PC to be procured locally,
* Windows b a registered trademark of Microsoft Corporation In the United States and/or other cc
* Intel b a registered trademark of Intel Corp. or Its affiliated companies h the United States snd/i
is of a PC are subject to change.
I Dimensions and weight
Items | Description (Width x Depth x Height, Weight) |
Main unit | 330 x 606 x 565mm, 63.0kg (manual stage) 330 x 630 x 565mm, 66.0kg (motor drive stage) |
Diaphragm pump | 145 x 256 x 217mm, 4.5kg |
I Optional Accessories
Energy Dispersive X-ray spectrometer (EDXJ
3-dimensional image display/measurement function 3D-VIEW
I Installation condition
Items | Description |
Room temperature | 15 to 30'C (At=±2.5'C/h or less) |
Humidity | 45%-70%RH |
Power source (TM3030) | Single-phase AC100 to 240V (Minimum: 90 [V], Maximum: 250 [V]) |
Grounding | 100 ohm or less |
♦Another power source for PC Is required,
■ Installation layout
•Power source fTMSOSO)
Single-phase AC100 to 240V, 50/60HZ 500VA
•Grounding
100 ohm or less
•Power source (PC)
Diaphragm t14_T
Pump 4 Power cables should be
• prepared by users
-®
Recommended table size: 1,200 x 800mm
Minimum table size: 1,000 x 800mm
Withstand load: 100kg or more
*A table with casters to not suitable to put a mail unltotTMSOSOcn.
* Recommended table size: 1,200 x 800mm, withstand load: 100tg or
♦Periodical maintenance Is required for this apparatus,
* Limited to Indoor operation.
•TM3030 is not approved as a medical device.
*Powercables, earth terminal and table should be prepared by users.
Notice: For correct operation, follow the instruction manual when using the instrument.
Specifications in this catalog are subject to change with or without notice, as Hitachi High-Technologies Corporation continues to develop the latest
technologies and products for our customers.
Copyright (C) Hitachi High-Technologies Corporation 2013 All rights reserved.
^Hitachi High-Technologies Corporation
Tokyo, Japan
http://wvw.hitachi-hitec.com/global/em/
24-14, Nishi-shimbashi,1-chome, Minato-kuTokyo,105-8717, Japan
zor technical consultation before purchase, please contact:contact@nst.hitachi-hitec.corr
HTD-E209 2013.5
Hitachi Tabletop Microscope
TM3030
Hitachi High-Tech
HITACHI
Gateway to Innovation
Tabletop
Microscope
"